NEW

Spacelabs Healthcare Partners with Masimo to Bring Advanced Measurements to Spacelabs Patient Monitors

Spacelabs Healthcare, a division of OSI Systems Inc., and Masimo announced that they have extended their partnership in patient monitoring to integrate SET pulse oximetry, NomoLine capnography, rainbow SET Pulse CO-Oximetry, SedLine brain function monitoring, and O3 regional oximetry into select Spacelabs multi-parameter monitors.

National Employer Groups to Hospitals and ASCs: Participate in the 2021 Leapfrog Surveys

Today, every hospital and ASC across the country received a letter signed by four influential national employer groups—the Health Transformation Alliance, ERISA Industry Committee, National Alliance of Health Care Purchaser Coalitions, and the Business Group on Health—requesting their participation in the annual Leapfrog Hospital Survey and Leapfrog Ambulatory Surgery Center (ASC) Survey.

Adapters to Test Light Cord Integrity

Healthmark Industries has introduced the Light Cord Adapters to its ProSys Instrument Care product line.

RSIP Vision Launches Advanced AI-Based Tool for Prostate MRI and Ultrasound Registration

RSIP Vision, an experienced leader in driving innovation for medical imaging through advanced AI and computer vision solutions, has announced a new prostate MRI-to-ultrasound registration tool.

Troubleshooting and Tech Support Tips For the Alaris 8000 Infusion System

Sponsored by Avante HS

By Wayne Mayberry, Biomedical Technician III

Wayne MayberryTo call for tech support or not to call for tech support? For many biomedical professionals, that is the question. While tech support will surely be able to help solve your issue, the convenience of being able to quickly fix your equipment on the spot is hard to ignore.

In this article, I will address two error codes for the Alaris 8000 Infusion System and show how they can be easily resolved with some experience and troubleshooting skills.

The “We Just Don’t Understand Each Other Anymore” Error

This error, signified with the code 200.5040, results when the two units, one being the 8015 Point-of-Care unit and the other one of the attached modules, cannot interface due to a difference in the firmware.

If you get this error, the first step is to determine what version of firmware the two units have by doing the following:

  1. Stop the error message by hitting Confirm.
  2. Press the Options key and the Page Down key.
  3. Press the Software key and the View key for each unit (PCU, PUMP, SYRINGE, etc.)

Contact the account of the unit you are working on to determine what version firmware the account is running. If any of the units do not match the version the hospital or account is running, then they must be flashed to match. In general, any 9.1x version firmware will run with any 9.1x or lower firmware, but a 9.3x may not run with a 9.1x. Software 9.3x must run with 9.3x, and so on.

The “What We Have Here is a Failure to Communicate” Error

This error code, 600.6835, is not listed in any of the recent service manuals but can be seen from time to time. The process for verifying this error is similar to the one above but requires a deeper dig into the software.

  1. Press Confirm to stop the alarm.
  2. Press the Options key.
  3. Press the Page Down key twice to get to page 2.
  4. Press the Network Status key and enter the passcode for the status.

If the wireless status is “disabled,” then the fix for this is fairly simple. If it is “enabled,” and you still get the error code, then you should call tech support for assistance.

  1. To fix this error, open the unit up in the system maintenance software and:
    • Run the “Clear Network Configuration” task
    • Run the “Transfer Network Configuration” task
    • Run the “Network Connectivity” task to verify connection.
  2. Turn the unit off and on to verify error has been cleared.
  3. If not, then please contact tech support for help.

The solutions for the two error codes we discussed today should help you quickly determine the problem with your Alaris 8000 system.

General Troubleshooting Tip: Take Detailed Notes

As always, be systematic and methodical when troubleshooting. Refining your troubleshooting skills to a razor-sharp edge will be a great asset to you over the course of your career. Take your time, don’t rush and confirm the information given after each step in the diagnosing process of the repair.

Always take notes when on the phone with tech support. Taking detailed notes is one of the most important ways to handle support issues quickly and efficiently. When performing a step or test sequence,having detailed notes takes the stress out of remembering the details of conversations and makes it easier to ensure the guidance given is followed in its entirety. Also, note-taking is a really good way to capture every detail for future use.

I hope this article helps in getting over some of the fear of contacting and asking for assistance from tech support. Remember they are there to help!

For more information on infusion pump sourcing, repair and parts, visit www.avantehs.com.

Wayne Mayberry is a Biomedical Technician at Avante Health Solutions Center of Excellence in San Clemente, California.